In the SMARTING Streamer, the signals are filtered with a high-pass filter for visualization purposes only, and this filter is always on. When recording or streaming the signals, the streamer stores and transmits the raw (non-filtered) data. The flat frequency response of the amplifier is 0-133 Hz when sampled at 500 Hz or 0-62.5 Hz when the sampling frequency is set to 250 Hz.